SOIL-DAQ™
TROUBLESHOOTING
Operator’s Manual EZ-DAQ™ version 1.4 24 of 25
TROUBLESHOOTING
1.0 Troubleshooting:
EZ-Daq panel shows values –but-
no values are on PC software.
No communication
between computer and
EZ-DAQ panel.
Network USB
connection not working.
Check IP address:
C:\DGSI\EZ-Daq\Config.ini
and EZ-DAQ panel must match.
Factory set for 10.11.12.42
Call DGSI for more details.
Belkin USB adaptor not working Drivers not installed Drivers located in EZ-DAQ CD
Under Belkin directory.
EZ-DAQ Panel unit on, but shows
no values in boxes.
Internal communication error.
Call factory.
EZ-DAQ panel shows unstable
values.
No sensor signal to
panel.
Unsteady A/D counts.
Sensor not plugged into back of
panel.
Sensor malfunction or cable
damaged.